Troubleshooting Guide
Error codes are shown on the main display for approx. 2 seconds. The program then
returns automatically to the previous mode.
Display Cause Solution
No segments appear No AC power is available Check the AC power supply
on the display
The power supply is not Plug in the power supply
plugged in
high The load exceeds the balance Unload the balance
capacity
low or Err 54 Something is touching Move the object that is
the weighing pan touching the weighing pan
err 54, typical Weighing system defect Contact Denver Service Center
App.err. Cannot store data: Increase load
Load on weighing pan too
light or no sample on pan
while application is active
dis.err. Data output not compatible Change the configuration in
with output format the operating menu
prt.err. Interface port for printer Reset the menu factory settings,
output is blocked or
Contact your local Denver
Service Center
err 02 Calibration parameter Calibrate only when zero
not met; e.g.: is displayed
balance not tared Press t to tare the balance
load on weighing pan Unload the balance
err 10 The t key is blocked when Press to clear the tare
there is data in the second tare memory and release the tare key
memory (net-total); only 1 tare
function can be used at a time
err 11 Tare memory not allowed Press t
The weight readout Unstable ambient conditions Set up the balance in another
changes constantly (excessive vibration or draft) area
at the place of installation
A foreign object is caught Remove the foreign object
between weighing pan and
balance housing
The weight readout The balance was not Calibrate/adjust the balance
is obviously wrong calibrated/adjusted
Balance not tared Tare or zero the balance
before weighing before weighing
If any other errors occur, contact your local Denver Service Center.