Block Calibration
(Select menu code: 1 9 7)
Blocks CAL KEY so calibration or adjustments cannot be made.
Only when menu access switch is in “locked” position (see page
13, “Menu Access Switch” for details).
Percent Weighing
Weighing in Percent (Menu Code 2 1 5)
Symbol displayed: %
This application program enables you to obtain weight readouts in percent
which are in proportion to a reference weight. The reference weight
readout is stored as a menu-defined percentage (factory setting: 100%).
1. Select Percent Code 2 1 5 (see back inside cover for instructions).
2. A “%” will be displayed on the bottom of the display.
3. Tare the balance and place the reference weight on the balance.
4. Press the FUNCTION key and the current reference percentage will
appear on the screen.
5. The screen will then flash and show the percentage on the balance.
6. Add or subtract parts and the percentage weight will be displayed.
Toggling between the Readout in Percent (%) and Weight (g)
After placing the sample on the weighing pan, you can toggle between the
readout in percent and the respective weight readout in grams by pressing
the FUNCTION key after you have stored the reference value.
Resetting the Reference Weight
1. Place the new reference weight on the balance.
2. Hold down CF for more than 2 seconds until the reading is in grams.
3. Press FUNCTION and “rEF (current setting) %” appears on display.
4. The screen will then flash and show the percentage on the balance.
5. Add or subtract parts and the percentage weight will be displayed.
7