Troubleshooting
Display Shows: Cause: Remedy:
No segments appear I/O key pressed Press the I/O key.
on the display. to turn OFF Display.
LL
The balance is Check the weighing pan
underloaded. for proper installation.
Weight is too light or there Increase reference
is no sample on balance. quantity.
HH
The load exceeds the Unload the balance or
balance capacity. look for obstruction.
Display capacity Decrease weight
is exceeded. on balance.
LLOO
Low battery. Replace battery.
EE --0022
The calibration operation Place balance in a
is being attempted and stable environment
vibration, air drafts or before calibrating.
other unstable conditions Before pushing the
exist. CAL/MODE button, the
LCD should display a stable
ZERO with the arrow mode.
EE --1100
There is a value stored Clear the memory or
in the memory when the turn the balance off to
TARE button is pressed. restore normal tare function.
EE --3300
An electronic malfunction Contact Denver Instrument
has occurred.
EE--5544
This indicates the Contact Denver Instrument
electronics are no longer
within the factory set
parameters. Objects have
dropped onto the weighing
pan or the balance itself
has been dropped.
The sensor is out of Contact Denver Instrument
specification. The balance
has been damaged by
rough handling. Liquids or
other materials have spilled
into the balance.
Do not attempt to access internal components or make any unau-
thorized repairs since this will void the warranty.